40 m Pure-Fiber HDMI for MRI Display Links | Phoossno

An MRI display integration project called for an HDMI link that could carry video over a 40 m (131 ft) cable run without creating a conductive path between the source and display. Phoossno developed a custom, separable pure-fiber HDMI assembly with a 40 m source-side run and a short display-side tail. The two assemblies mate at an optical interface, so the main cable route and the clearance behind the display can be specified independently.

Electrical Isolation in an HDMI Link

At the source end, the active HDMI head converts the electrical signal to light. The receiver at the display end converts it back to an electrical HDMI signal.

A fiber HDMI label describes the transmission medium. Some active optical cables (AOCs) still use copper conductors for power or auxiliary signals. When galvanic isolation is a design requirement, specify that no continuous conductive path may cross the isolation boundary through the cable assembly. Review the power and grounding arrangements at both ends, along with any other connections that could bridge that boundary. Phoossno medical imaging connectivity solutions

40 m Assembly Configuration

Assembly

Configuration

Installation purpose

Source-side assembly

HDMI source head with a nominal 40 m run of four-fiber OM3 multimode cable

Covers the main cable route

Display-side assembly

HDMI display head with a 0.5 m or 1 m tail

Fits the available space and service clearance behind the display

The 40 m dimension applies to the source-side assembly. If the project specifies an overall source-to-display length, account for the selected display tail and the slack required at each termination.

For a route that crosses the room’s RF shield, check the dimensions of the optical termination that will actually pass through the opening. At UC San Diego’s Center for Functional MRI, for example, fiber can pass through a waveguide in the penetration panel, while current-carrying cables use filtered feedthroughs. Confirm the permitted route and available clearance with the project’s RF shielding contractor or site team. UC San Diego 3T penetration panel and waveguide guidance

Specifications to Confirm Before Ordering

· Cable route and termination clearance. Supply the measured route length, any feedthrough restrictions, and the preferred display-tail length and connector orientation.

· Video format and endpoint power. Provide the source and display model numbers, required resolution and refresh rate, and any EDID or HDCP requirements. Confirm how each active HDMI head will be powered.

· Isolation boundary and component location. Identify where the conductive path must be broken and which components will be installed in the scanner room. The project team should verify the MR safety status of those components for their intended locations and conditions of use. FDA-recognized standard for MR safety marking

Need a custom fiber HDMI assembly for an MRI display project? Send Phoossno your device models, target video format, measured cable route, isolation boundary, display-side clearance, and required quantity. Discuss your project with Phoossno